Analysis of VOCs in Liquids through Vaporization in a Tubular Oven Monitored by Chemical Ionization Mass Spectrometry

Taous Abar,Hélène Mestdagh,Michel Heninger,Joel Lemaire
DOI: https://doi.org/10.3390/s24041048
IF: 3.9
2024-02-07
Sensors
Abstract:The analysis of chemical compounds present at trace levels in liquids is important not only for environmental measurements but also, for example, in the health sector. The reference technique for the analysis of Volatile Organic Compounds (VOCs) in liquids is GC, which is difficult to use with an aqueous matrix. In this work, we present an alternative technique to GC to analyze VOCs in water. A tubular oven is used to completely vaporize the liquid sample deposited on a gauze. The oven is heated in the presence of a dinitrogen flow, and the gas is analyzed at the exit of the oven by a chemical ionization mass spectrometer developed in our laboratory. It is a low magnetic field Fourier Transform Ion Cyclotron Resonance (FT-ICR) optimized for real-time analysis. The Proton Transfer Reaction (PTR) used during the Chemical Ionization event results in the selective ionization of the VOCs present in the gas phase. The optimization of the desorption conditions is described for the main operating parameters: temperature ramp, liquid quantity, and nitrogen flow. Their influence is studied using a 100 ppmv aqueous toluene solution. The analytical method is then tested on a mixture of seven VOCs.
engineering, electrical & electronic,chemistry, analytical,instruments & instrumentation
What problem does this paper attempt to address?
The main problem that this paper attempts to solve is to develop a new technique for analyzing volatile organic compounds (VOCs) in liquids. Specifically, this technique aims to address the difficulties of gas chromatography (GC) in handling aqueous samples, such as the injection of water may damage the stationary phase of the column, and a long time is required for chromatographic separation. In addition, this technique has been optimized for cases where the sample amount is limited or contains a low - concentration VOC mixture. The long - term goal is to use this technique to analyze VOCs in sweat and identify sets of pathological biomarkers. To achieve this goal, the authors proposed an alternative method to GC, that is, vaporizing liquid samples through a tube furnace and using chemical ionization mass spectrometry (CI - MS) for real - time monitoring. This method uses a low - magnetic - field Fourier transform ion cyclotron resonance mass spectrometer (FT - ICR MS) and optimizes the desorption conditions, including the main operating parameters such as temperature gradient, liquid volume, and nitrogen flow rate. Through these optimizations, this method can effectively analyze a mixture of seven VOCs in aqueous solutions, showing good selectivity and sensitivity. In conclusion, this study provides a new method for VOC analysis in complex water - based samples that is efficient, fast, and applicable, and has important application values in fields such as environmental monitoring and health diagnosis.