Therapeutic effect of domestic ocduder on patients with congenital heart diseases

WANG Zhen,XIE Qilian,ZHANG Milin,GAO LEi,TAN Huilian,LIU Ling
DOI: https://doi.org/10.3969/j.issn.1001-1439.2007.05.004
2007-01-01
Journal of Clinical Cardiology
Abstract:Objective:To evaluate the therapeutic effect of domestic occluder device on patients with congenital heart diseases. Method:Six hundred and sixty-two patients were treated with the occluder devices through angiography and transthoracic echocardiography(TTE). Result:The atrial septal defect (ASD) diameter was (16.8±7.6)mm by TTE. The diameter of the occluder selected was (20.3±8.8)mm. The ventricular septal defect (VSD) diameter was (6.7±3.4)mm measured by left ventricular angiography and the diameter of occluder selected was(8.9±4.2)mm. The narrowest diameter of the patent ductus arteriosus(PDA) was (6.6±3.2)mm by angiography. The diameter of occluder selected was (12.7±4.4)mm in aortic side. The successful rate was 95.8% . No residual shunts were found in all successful cases after the procedure. And after the percutaneous balloon pulmonary valvuloplasty (PBPV), the systolic pressure gradient across the pulmonary valve all decreased to lower than 2.66 kPa.Conclusion:The therapy of congenital heart diseases with domestic occluder is safe and reliable and has high achievement ratio.