Electrochemical Impedance Method Based On Enzyme-Catalyzed Depositing Enlargement Used For The Immunoassay Of Human Immunoglobulin M

Zhou Guzhen,Li Jishan,Jiang Jianhui,Shen Guoli,Yu Ruqin
DOI: https://doi.org/10.3321/j.issn:0253-3820.2006.02.003
IF: 1.193
2006-01-01
Chinese Journal of Analytical Chemistry
Abstract:A novel electrochemical immunoassay method based on enzyme-catalyzed depositing enlargement is proposed. Primarily, the horseradish peroxidase (HRP)-labeled antibody (goat-anti-immunoglobulin M (IgM)-HRP) was immobilized on the electrode surface by sandwich method (antibody/antigen/antibody-HRP). Then the undissolving deposition was formed on the electrode surface by using HRP catalyze hydrogen peroxide to oxidize 3,3'-diaminobenxidine tetrahydrochloride dihydrate. So the electrochemical signal of the electrode is enlarged greatly. The immobilization of antibody was realized by using protein A. The effects of experimental conditions, such as the concentration of antibody, pH of incubated solution, the amount of enzyme-labeled antibody and the depositing time on the performances of immunassay were discussed respectively. The formed immunosensor can be used to quantitatively determine human IgM in the range of 2.1 similar to 670 mu g/L, and the detection limit is 0.08 mu g/L.
What problem does this paper attempt to address?