Application of SENSE technique and diffusion sensitivity coefficient(b-value)in MR diffusion weighted imaging of rabbit VX2 liver tumors

Liqiang Zheng,Xinqing Jiang,Qi Xie,Shamei Bai,Hongzhen Wu,Mei Wu
DOI: https://doi.org/10.3969/j.issn.1009-6817.2008.01.001
2008-01-01
Abstract:Objective To investigate the application of SENSE technique and diffusion sensitivity coefficient(b-value)in TSE MR diffusion weighted imaging(DWI)of experimental rabbit VX2 liver tumors.Methods TSE MR DWI scans were performed in all the 24 New zealand rabbits at 2-3 week after VX2 liver tumor implantation.The reduction factor of imaging parameter was taken as 1 and 2,the shortening of scanning times and the changes of imaging signal-to-noise ratio(SNR)were observed by using SENSE technique.The b-value was used by total 6 values from 100 s/mm2 to 600 s/mm2.SNR and apparent diffusion coefficient(ADC)between VX2 liver tumor with different b-values and normal liver tissue were measured for each rabbit and statistically analyzed.Results When use reduction factor 2,the scanning time was shortened by 13.3%,while the imaging SNR had no difference during DWI.When the b-value was 400 s/mm2,the rabbit VX2 liver tumors had a highest SNR(115.5±71.6);ADC value was 1.575±0.218×10-3mm2/s(P< 0.05).Conclusion Reasonably use SENSE technique can shorten the scanning time but the imaging SNR has no difference.The b-value teken as 400 s/mm2 is optimal,in which the SNR between VX2 liver tumor and normal liver tissue is highest and the ADC value is also most accurate.