Diffusion weighted imaging of transplanted kidney at 3.0 T MR scanner

SHAN Ren-feng,PENG Gui-zhu,ZHOU Fu-qing,XIAO Guo-sheng,RAO Xue-feng,WU Bo
DOI: https://doi.org/10.3969/j.issn.1005-8982.2010.02.025
2010-01-01
Abstract:[Objective]To evaluate the feasibility of diffusion weighted imaging(DWI)of transplanted kidney at 3.0T MR scanner and the inter-observer and intra-observer reproducibility of apparent diffusion coefficient(ADC)value.[Methods]Transplanted kidney T1WI,T2WI and DWI imaging was performed on 23 cases with transplanted kidney.Two doctors evaluated the quality of T1WI,T2WI and DWI images with 4 grades criterion.Bilateral kidney SNRs of cortex and medulla were measured on T1WI,T2WI and DWI images by one doctor and reconstructed on workstation,two doctors measured the ADC values of cortex,medulla and total ADC(cADC,mADC and totADC)respectively and they were measured again in 10 cases by one of the two doctors 2 weeks later.The values were statistically analyzed.[Results]The SNRs of transplanted kidney could meet the request of imaging.According to both observers,95.6%(22/23)of T1WI and 91.3%(21/23)of T2WI images were scored 4,82.6%(19/12)and 95.6% (22/23) of DWI images were scored 4 and 3 respectively.Analysis showed that the intra-observer and inter-observer reproducibility of ADC value(cADC,mADC and totADC)Was good without significant difference.[Conclusion]The inter-observer and intra-observer reproducibility of ADC values of transplanted kidney cortex.medulla and total ADC measured at 3.0T MR is high,and it is feasible to perform functional MRI of rat kidney at 3.0T MR scanner.