Blood Oxygen Level-Dependent MR Imaging of Rat Kidney at 3.0T MR Scanner: Initial Experience

YANG Xue-dong,CAO Ju,WANG Xiao-ying,LI Xiao-mei,XU Yu-feng,JIANG Xue-xiang
DOI: https://doi.org/10.3321/j.issn:1003-3289.2007.06.008
2007-01-01
Abstract:Objective To evaluate the feasibility of BOLD MR imaging of rat kidney at 3.0T MR scanner and the interobserver and intraobserver reproducibility of R2 value. Methods Twelve male healthy Wistar rats were performed kidney T2WI and BOLD imaging. Two doctors evaluated the quality of the T2WI and BOLD images with 5 grades criterion. Bilateral kidney SNRs of cortex and medulla were measured on T2WI. BOLD images were reconstructed on workstation, two doctors measured the R2 values of cortex and medulla respectively and one of the two doctors measured them again 2 weeks later. The values were statistically analyzed. Results 91.7%(11/12) vs 83.3%(10/12) of T2W images scored 5 grades (Kappa value=0.625, P=0.02), and 83.3% (10/12) vs 91.7% (11/12) of BOLD images scored 4 grades and above (Kappa value=0.586,P=0.008) by the two observers respectively. Bland-Altman analysis showed that the intraobserver and interobserver reproducibility of R2 value was good. The correlation coefficient was: interobserver 0.96 and 0.98, intraobserver 0.97 and 0.98 for cortex and medulla respectively. Coefficient of variation between rats was 17.80% and 17.96% in cortex and medulla, respectively. Conclusion The interobserver and intraobserver reproducibility of R2 value of rat renal cortex and medulla measured at 3.0T MR is high, and it is feasible to perform functional MRI of rat kidney at 3.0T MR scanner.