Effect of Shared Cavity on Electromechanical Performance of Piezoelectric Based Micro-machined Ultrasonic Transducer Array.
Weili Wang,Yihsiang Chiu,Dan Gong,Shenglin Ma,Wen Lei,Hungping Lee,Huguang Liu,Yufeng Jin
DOI: https://doi.org/10.1109/nems.2019.8915657
2019-01-01
Abstract:In this paper, we present a 2μm-thick aluminum nitride (AlN) based piezoelectric micro-machined ultrasonic transducer (PMUT) array, these PMUTs share one cavity. The Finite Element Model (FEM) method is used to perform various electromechanical performances of the PMUT array, such as resonance frequency, displacement frequency response, electromechanical coupling coefficient and sound pressure output to evaluate the time and frequency domain performance of the PMUT array. The results are verified by an impedance analyzer and hydrophone. The results show that the PMUT's resonance frequencies are from 106kHz to 152 kHz, so it can be used for distance detection. The sound pressure of a single PMUT received by a hydrophone is from 0.208Pa to 0.752Pa, and when these PMUTs are simultaneously emitted, the sound pressure is 2.57Pa, which is 3-12 times larger than single PMUT. At a fixed resonant frequency, the sound pressure intensity can be controlled by changing the number of transmitted PMUTs to control the ranging range and accuracy.