Determination of Cationic Surfactant by Flow Injection and Ionic Selective Electrode Analysis

WEI Fu-xiang,LI Tao,WANG Rui
DOI: https://doi.org/10.3969/j.issn.1004-4957.2009.02.021
2009-01-01
Abstract:A on-line flow injection and ionic selective electrode(FIA-ISE) method was developed for determination of a cationic surfactant,cetyltrimethylammonium bromide(CTMAB),in water by using a self-made tubular polymer sensitivity membrane electrode as detector,connected with flow injection and potentiometer.The detection of CTMAB was performed by using 340 μL sampling volume and 15 cm length of disperse coil,and employing double distilled water as carrier stream with a flow rate of 1.5 mL·min-1.The electrode showed Nernst response for CTMAB over the concentration range of 6.4×10-7-1.6×10-4 mol·L-1 with a correlation coefficient of 0.999 2 at a slope of-53.07 mV/pcCTMAB.The spiked average recovery was near 100%.The method was simple,rapid,and could be applied in the determination of CTMAB in water sample.
What problem does this paper attempt to address?