Determination of Gaseous Semi- and Low-Volatile Organic Halogen Compounds by Barrier-Discharge Atomic Emission Spectrometry

Yifei Sun,Nobuhisa Watanabe,Wei Wang,Tianle Zhu
DOI: https://doi.org/10.1016/s1001-0742(12)60032-1
2013-01-01
Abstract:A group parameter approach using “total organic halogen” is effective for monitoring gaseous organic halogen compounds, including fluorine, chlorine, and bromine compounds, generated from combustion. We described the use of barrier-discharge radiofrequency-helium-plasma/atomic emission spectrometry, for the detection of semi– and low-volatile organic halogen compounds (SLVOXs), which can be collected by Carbotrap™ adsorbents and analyzed using thermal desorption. The optimal carrier gas flow rates at the injection and desorption lines were established to be 100 mL/min. The detection range for SLVOXs in the gaseous samples was from 10 ng to tens of micrograms. Measuring F was more difficult than measuring Cl or Br, because the wavelength of F is close to that of air. The barrier-discharge radiofrequency-helium-plasma/atomic emission spectrometry measured from 85% to 103% of the SLVOXs in the gas sample. It has been found that Carbotrap B is appropriate for high-boiling-point compounds, and Carbotrap C is suitable for the determination of organic halogen compounds with lower boiling points, in the range 200–230°C. Under optimal analysis conditions, a chlorine-containing plastic was destroyed using different oxygen concentrations. Lower oxygen concentrations resulted in the production of lower amounts of organic halogen compounds.
What problem does this paper attempt to address?