PERFORMANCE TEST ON BD-PND NEUTRON BUBBLE DETECTOR

Liang Shiqiang,Yu Tao,Qiao Haitao,He Lihua,Luo Hailong,Wang Zhiqiang
DOI: https://doi.org/10.3321/j.issn:1000-8187.2006.05.006
2006-01-01
Abstract:The performances(homogeneity,linearity,response,reproducibility and angular response,etc.) are tested about a batch of calibrated BD-PND bubble detectors which have not worked for two years after receipt according to GB/T 14959 and the performance requirements of ISO/DIS 21909.The results show that the technical performances of the BD-PND bubble detectors basically meet the performance requirements of ISO/DIS 21909,when corrective response coefficients of BD-PND bubble detectors are used.They can be used to measure personal neutron dose equivalent in the low neutron dose equivalent level.However,these BD-PND bubble detectors would cautiously be used because they will reach the end of their life span.