Instrument Performance of A Radon Measuring System with the Alpha-Track Detection Technique

S Tokonami,W Zhuo,H Ryuo,H Yonehara,Y Yamada,M Shimo
DOI: https://doi.org/10.1093/oxfordjournals.rpd.a006118
2003-01-01
Radiation Protection Dosimetry
Abstract:An instrument performance test has been carried out for a radon measuring system made in Hungary. The system measures radon using the alpha-track detection technique. It consists of three parts: the passive detector, the etching unit and the evaluation unit. A CR-39 detector is used as the radiation detector. Alpha-track reading and data analysis are carried out after chemical etching. The following subjects were examined in the present study: (1) radon sensitivity, (2) performance of etching and evaluation processes and (3) thoron sensitivity. The radon sensitivity of 6.9x10(-4) mm(-2) (Bq m(-3) d)(-1) was acceptable for practical application. The thoron sensitivity was estimated to be as low as 3.3x10(-5) mm(-2) (Bq m(-3) d)(-1) from the experimental study.
What problem does this paper attempt to address?