Experience with transcatheter closure of secundum atrial septal defects using the China-made septal occluder: a single centre study in 461 consecutive patients

Dong-jiang MA,Yu-shun ZHANG,Huan LI,Zheng-xue DAI,Dong-liang WU,Xu-mei HE
DOI: https://doi.org/10.3760/cma.j.issn.1673-4904.2006.30.001
2006-01-01
Abstract:Objective To evaluate the safety and efficiency of transcatheter closure of secundum atrial septal defects (ASD) with the China-made septal occluder. Methods There were 461 patients,all patients were occluded with the China-made septal occluder including 176 patients with deficient or insufficient residual rim, 22 multiple ASDs, 21 ASDs with atrial septal aneurysm, 37 with large ASD, 2 residual fistula after surgery repair of ASD and 32 combined congenital heart disease. The closure procedure was guided by fluoroscopy and transthoracic echocardiography(TTE)or transescophageal echocardiography (TEE). For patients with ASD and other deformities, ASD was occluded after the other deformities were treated. TTE was performed immediately and after the procedure to evaluate the efficiency. Results The defected diameter of ASD was(26.4±12.6)mm .The diameter of the occluder selected was(28.9±8.0) mm . The total of 465 septal occluders had been implanted. The incidence of serious complication was 1.3%. And the other complications were 62 cases. Conclusion The incidence of serious complication of the China-made septal occluder is very low, which suggests that transcatheter occlusion of atrial septal defects with the China-made occluder is a safe and effective method.
What problem does this paper attempt to address?