Determination of Glycidyl Butyrate by Gas Chromatography

HUANG Qiang,MENG Qing-Yi,BAN Chun-Lan,ZHANG Rui
DOI: https://doi.org/10.3969/j.issn.1004-8138.2011.05.072
2011-01-01
Abstract:Glycidyl butyrate of epichlorohydrin reaction system was detected by quick analysis of gas chromatography.The SE-30(Φ3mm×1.5mm)column was used with the gas flow of 48.86mL/min.The column temperature was 100℃ with the gas room temperature of 185℃,and the thermal conductivity detector of 185℃.There was a good linear relationship between the peakarea(y) and the quantity of the glycidyl butyrate(x) with the regression equation of y=1594.1x+1370.1(R2=0.9993,n=6),and limit of detection(S/N=3) was 0.1778mg/mL with the precision of 0.19% and the accuracy of 0.0028%.The recovery was 96.8%—103.6% with average recovery of 99.3%.This method is easy,accurate and useful.
What problem does this paper attempt to address?