Ultrasonic tissue lesion monitoring in HIFU hyperthermia

HU Qi-long,LI Wei,TAO Kan,XIAO Xiang-lin,NIU Jin-hai
DOI: https://doi.org/10.3969/j.issn.1000-3630.2010.03.009
2010-01-01
Abstract:An experimental study of ultrasonic tissue lesion monitoring has been conducted.The high intensity focused ultrasound(HIFU) of specific power and frequency is focused on porcine tissue in vitro,and meanwhile the B-scan ul-trasonic images of the focusing area are acquired and processed to find out the relationship between the image change and the tissue lesion and a method for monitoring the tissue lesion with B-scan ultrasonic image is provided.In the si-mulative experiment,M-mode signals are acquired and temperatures measured by probes,and the correlation between them are studied.It is proved that after HIFU irradiation the tissue in the focusing area degenerates and the gray level rises significantly,and that more signal changes are found in M-mode than B-scan.Compared with the former methods,the experimental system could be a new visible real-time one to monitor the tissue lesion in HIFU hyperthermia.