The Ultrasound Fields Estimation Using Uncooled Infrared System

Yu Ying,Shen Guofeng,Bai Jingfeng,Chen Yazhu
DOI: https://doi.org/10.1063/1.4757379
2012-01-01
AIP Conference Proceedings
Abstract:The ultrasound fields through a certain thickness of tissue are a critical parameter in the high intensity focused ultrasound (HIFU) treatment. Although the ultrasonic fields in tissue are difficult to measure, the temperature elevation induced by ultrasound can be obtained to estimate the ultrasonic intensity. In this work, the ex-vivo experiments were conducted in porcine muscle with the thickness of 10-30 mm. The distance from the transducer to the surface of muscle can be varied. The temperature distribution in tissue-air surface can be obtained from an uncooled infrared system (spatial and thermal resolutions are 0.59 mm and 0.08 degrees C). According the Snell's law, the intensity at interface is about four times of the incident wave when ultrasound wave propagates from tissue to air, due to the total internal reflection. Then the intensity distribution of incident wave can be estimated by temperature distribution in tissue-air surface. The experimental temperature was consistent with that of simulations. The maximum temperature in the surface was 46.27 degrees C heated for 5s, and the maximum estimated peak incident sound intensity was 458.2 W/cm(2). This work may offer implications and information for treatment planning toward optimizing focused ultrasound surgery in deep tumor.
What problem does this paper attempt to address?