Calibration Analysis of H-ADCP Automatic Flow Measurement System in Wangting (Interchange) Hydrological Station

任晓东,隆威,唐晓春,耿欣欣
DOI: https://doi.org/10.3969/j.issn.1674-9405.2013.02.017
2013-01-01
Abstract:Through comparative analysis of H-ADCP measurement data and manually measured flow data in Wangting (Interchange) hydrological station, combining with the characteristics of H-ADCP, the index velocity method is used to calibrate the relation of cross-section velocities which are measured by H-ADCP and cableway current meter method. The steady line precision meets the specification requirements. On this basis, the measured data are compared for error statistics and velocity formula is given. The online measuring data meet the practical requirements.