Full Peak Fitting Function to Γ-Ray Spectra from a Quick NaI Whole Body Counter

齐荣,毛永,陈熙萌
DOI: https://doi.org/10.3321/j.issn:0253-3219.2008.05.003
2008-01-01
Nuclear Techniques
Abstract:Gamma ray spectra were measured by a Φ8"×4"NaI(Tl) detector with different radioisotope sources to test a full-energy peak fitting function,which is a Gaussian plus a high-energy tail correction and a low-energy cor-rection.The fitting function can properly describe a γ-ray peak with nine parameters,which are automatically opti-mized.The parameters for 57Co,137Cs,60Co and 40K and the corresponding analysis were given.Considering requests on analysis time and credibility of the γ spectra by a NaI whole body counter,the function was simplified on the basis of experiment accuracy.The high-energy tail correction term can be ignored due to lower contribution,the peak ad-dress and FWHM of the Gaussian,and the low-energy tail correction term,could be bonded together to reduce the freedom to four parameters in the fitting.The results show that the Analytic Improved Figure-of-Merit(AIFOM) of the fitting for γ-ray peaks of 57Co,137Cs,60Co and 40K measured by the NaI detector are less than 0.01%.