Low-dose Technique of Lung Screening with a 16-slice CT Scanner

Zhang Jun,Xiao Xiangsheng,Liu Shiyuan
DOI: https://doi.org/10.3969/j.issn.1006-5741.2006.03.007
2006-01-01
Abstract:To explore the low dose related with the body build in lung screening using a 16-slice CT scanner. Materials and Methods: Lung screening was performed in 150 healthy volunteers(male 89,female 61,height 152~182cm, mean 166cm; weight 45~86kg, mean 63.1kg ) during Mar 2004 to Apr 2005 with a TOSHIBA Aquilion 16-slice CT scanner. The imaging parameters were 1.0 mm collimation, p=0.938, 120kV, 0.5s/Rot, and FOV 280~330mm. Slice thickness of 7.0 mm with lung window and 10mm with mediastinal window setting were reconstructed, respectively. Four dosages were used and compared in screening.Results: All patients had obtained the good quality images using a dose of no more than 20 mAs. Most of them only need a dosage of 10mAs(96 patients, 64%), and the others need 5mAs(16 patients,10%) or 20mAs(38 patients,26%). The 95%confidence interval of the ratio of height and weight was 1.00~1.03 for 10mAs, 0.83~0.88 for 5mAs, and 1.13~1.22 for 20mAs. Conclusion: The lowest screening dose could change with the body build. The dosage of 10mAs might be applied in most screening, and that of 20mAs almost in all of patients.