EXPERIMENTALLY STUDY OF A TEMPERATURE SENSOR WITH CAPACITORS

Sun Yue,Sun Dong,Li Ling
DOI: https://doi.org/10.3969/j.issn.1000-3258.2007.03.019
2007-01-01
Abstract:From 77 K to 300 K, the relationships of capacitance vs temperature of some commercial solid medium capacitors, like tantalum, terylene, normal ceramic, glass-glazed and integrative monolithic ceramic, are measured, respectively. The regressive results indicate that, the monolithic ceramic capacitor presents a good linear response, sensitivity, and standard deviation, i.e. -10pF/℃, 0.1 ℃, 0.016; while the terylene does a similar characteristic, about +17pF/℃, 0.06 ℃ and 0.01, respectively, within the positive and negative temperature region. Moreover, the relations of capacitors mentioned above is almost unacted on an external magnetic field, even its intensity is up to 0.5 tesla, and whether tested temperature is positive or negative. Therefore, the works would be more beneficial to develop an anti-electromagnetic thermoscope with a negligible self-heating effect and conduct thermal coefficient at low temperature.
What problem does this paper attempt to address?