Research on measurement of anterior chamber parameters by Pentacam in cataract patients

JIANG Dan,LIU Xin-Quan,WU Qiang,FANG Jian,LU Bin
DOI: https://doi.org/10.13389/j.cnki.rao.2010.08.015
2010-01-01
Abstract:Objective To investigate the differences in anterior chamber depth(ACD)and anterior chamber angle(ACA)measured by Pentacam and UBM,their differences with ACD by A-scan and correlation with lens density in age-related nuclear cataract.Methods ACD and ACA were measured in 52 cases(52 eyes)with age-related nuclear cataract by Pentacam and UBM,respectively.Pentacam and A-scan were used to measure ACD after mydriasis in patients.Results ACD values were(2.63±0.39)mm by Pentacam and(2.63±0.34)mm by UBM,there was no statistical difference(P=0.595,r=0.969);ACA values were(31.27±5.46)° by Pentacam and(30.89±5.37)° by UBM,there was no difference(P=0.218).Significant correlations were found in ACD and average ACA by Pentacam and UBM(r=0.969,r=0.759).After mydriasis,ACD values were(2.59±0.21)mm by Pentacam and(2.79±0.58)mm by A-scan,there was statistical difference(P<0.05),and had correlation with lens density(r=0.186,P<0.05).When the lens density were less than 20,from 20 to 40 and more than 40,the correlations for difference of ACD between Pentacam and A-scan and lens density was 0.079,0.264 and 0.288(all P<0.05).Conclusions Pentacam can be a good choice for measuring ACD and average ACA in patients with age-related nuclear cataract,and it should be used flexibly and analyzed synthetically in clinical works.