Mechanical Test of Impaction Bone Grafting in Revision of Tibia (Initial Stability of Short and Long Stem)

Li Feng,Lou Siquan,Zhang Ke
DOI: https://doi.org/10.3969/j.issn.1672-9935.2006.08.008
2006-01-01
Abstract:Objective To investigate the effect of the type of stem and the graft impaction on the initial stability of the tray in revision total knee arthroplasty through impaction bone grafting.Methods Ten tibia from 5 cadavers were randomizedly divided into two groups. The T3 type (AORI classification) contained bone defect was made. The trays with short (35 mm) or long (75 mm) stems were randomizedly implanted in the tibia on both sides from the same cadaver using morsellised impaction bone grafting. Bone mineral density around the implant and X-ray image were observed. The biomechanical tests were performed under circular load (1 000 N).Results There was no statistically significant difference in the micromotion between the two types of stems. The migration of stem was correlated inversely with the density of the impacted graft.Conclusion The impaction bone grafting can be employed in the severe bone defect in tibia. If support of the intact cortex and the bone cement is sufficient, both short and long stems can achieve adequate initial stability.