Finite Element Modeling of Ossicle for Sound Transmission

黄新生,姚文娟,刘骏帧,付黎杰,林江
DOI: https://doi.org/10.3969/j.issn.1008-6358.2008.02.041
2008-01-01
Abstract:Objective:To evaluate the hearing improvement following partially ossicular replacement with different materials. Methods: The CT data of normal ossicle, were digitalized and imported into PATRAN( a world class commercial software. ) to build a 3 dimension finite element model. Then the ossicle model was analyzed based on dynamic transmitting vibration principles and the calculation results were evaluated and compared with various material preferences. Results:The ossicle 3D model established in the present study was in accord with biological structure of the human, the frequency reaction curves obtained by the model proposed in this study agreeed well with the fresh temporal bone ones. This research demonstrated that,with the increasing of elastic modulus,the amplitude of stapes footplate would also increase and get to the amplitude of stapes footplate more closely. After the elastic modulus getting bigger than 115000 gPa,the change rate of amplitude curve would be relatively small,which was almost a horizontal line. Conclusion:The model in this study is nicety. In stead of paying mainly attention to the material density when choosing the artificial replaced body materials in clinic,the physical preferences of the chosen material should focus mainly on the Young's modulus,when the elastic modulus increases so does,the hearing recovering effect,while as the elastic modulus comes to 115000 gpa,which means the elastic modulus arrive at the limit value for the artificial body, the hearing recovering effect no longer gets better.
What problem does this paper attempt to address?