Finite Element Analysis the Ring Diameter Ofjunction Between Stapes Protheses and Incus Long Process on Sound Transmission in the Middle Ear after Stapes Replacement

谢晓凤,黄新生,姚文娟,王建中,付黎杰,刘骏帧
DOI: https://doi.org/10.3969/j.issn.1008-6358.2008.03.046
2008-01-01
Abstract:Objective:To evaluate the ring diameter of junction between stapes protheses and incus long process on sound transmission in the middle ear after stapes replacement,and to use this information to guide stapes surgery.Methods:The CT data of ossicle from Zhongshan Hospital of Fudan University,were digitalized and imported into PATRAN(a world class commercial software.) to build a 3 dimension finite element model(FEM).Then the ossicle model was analyzed based on dynamic transmitting vibration principles and the calculation results were evaluated and compared with results from examination。Using stapes protheses of different ring diameter to reconstruct Ossicle chain in stapes replacement and analyzed based on dynamic transmitting vibration principles,the calculation results were evaluated and compared with natural Ossicle chain.Results:The ring diameter of junction between stapes protheses and incus long process can effect the amplitude of protheses in ring stapes replacement.The smaller the ring diameter,the amplitude of protheses is closer to the normal,and the hearing recovering effect will be better.Conclusion:The ring diameter of junction between stapes protheses and incus long process can effect the hearing recovering after stapes replacement.
What problem does this paper attempt to address?