Abstract:Objective To investigate the effect of managed ventricular pacing(MVP) function dual chamber pacemaker on sick sinus syndrome(SSS) patients with a high proportion of right ventricular pacing.Methods A total of 35 cases of SSS patients with normal atrioventricular conduction or intermittent atrioventricular block were implanted with dual chamber pacemaker(DDD/R)at least two years ago.They all wanted to change DDD/R into MVP function pacemaker instead.Before the operation,the proportion of ventricular pacing of DDD/R was over 40% more than a month.Program control of MVP was randomly on(MVP-on group,n=17) or off(MVP-off group,n=18).The brain natriuretic peptide(BNP) level in plasma of all patients were determined before the operation and 6 monthes after the operation.Left ventrieular end diastolic diameter(LVEDD),left ventrieular end systolic diameter(LVESD) and left ventricular ejection fraction(LVEF) were also determined.Six monthes later after the operation,program control was applied to obtain the proportion of right ventricular pacing and the events of atrial high-frequency.Results Before the operation,there was no statistical different between the MVP-on group and the MVP-off group in plasma levels of BNP,LVEDV,LVESV and LVEF(P>0.05).After the operation,the parameters of echocardiography between the MVP-on group and the MVP-off group didn′t have any statistical different either(P>0.05).In MVP-on group,the plasma BNP levels were significant decreased after the operation(P<0.05).Compared with the MVP-off group,in the MVP-on group,the proportion of right ventricular pacing and the events of atrial high-frequency were significant decreased(2.6%±0.6% vs 48.4%±10.5%,15.7%±2.3% vs 62.0%±7.8%;P﹤0.01).Conclusions DDD/R pacemaker with MVP function could decrease the proportion of right ventricular pacing and the occurrence of atrial high-frequency events in the SSS patients with high proportion of right ventricular pacing.
What problem does this paper attempt to address?