Design of a selective interface for electrospray ionization-mass spectrometry.

Chen Huanwen,Zhou Yueming,Wang Zhichang,Luo Mingbiao,Zhang Xie
DOI: https://doi.org/10.3969/j.issn.1001-232X.2006.02.004
2006-01-01
Abstract:A novel electrospray ionization-mass spetrometry (ESI-MS) interface with the ability to discriminate ions of different momentum and mobility, and neutral particles was designed and its performance was investigated. Because the interface is able to considerably increase the transmission efficiency of single-charged ions and reduce the probability of neutral molecuels and multiple-charged ions to enter the vacuum system, the chemical noise is reduced significantly and the S/N ratio is improved. The interface is helpful to keep the system clean and stable, and is hopeful to be used in nano-MS, HPLC-MS and other miniaturized mass spectrometers.
What problem does this paper attempt to address?