Evaluation of the US2025A urine sediment analyzer in clinical application

YI Lin,XU Na,LIU Ran
DOI: https://doi.org/10.3969/j.issn.1672-9455.2010.10.014
2010-01-01
Abstract:Objective To evaluate the performance of US2025A urine sediment analyzer by comparing the results of US2025A with that of manual microscopy.Methods 1 000 urine samples were inspected by US2025A urine sediment analyzer and microscope method simultaneously.Results Compared with manual microscopy,the agreement rate of detection of RBC,WBC and casts was 84%,93.6% and 88.4%,respectively;The sensitivity of RBC,WBC and casts was 91.1%,91.5% and 85.7%,respectively;The specificity of RBC,WBC and casts was 82.1%,94.3% and 89.4%,respectively;The false positive rate of RBC,WBC and casts was 17.9%,5.7% and 10.6%,respectively;The false negative rate of RBC,WBC and casts was 8.9%,8.5% and 14.3%,respectively.Conclusion US2025A urine sediment analyzer can be used for the analysis of the quantitative urine sediment in clinic.However,the results are likely to be influenced by the factors such as crystals,bacteria,epithelia,and etc.As a result,the US2025A analyzer still cannot substitute manual microscopy in the urinary sediment inspection.