Bone Responses to Titanium Implants Surface-Roughened by Sandblasted and Double Etched Treatments in a Rabbit Model.

Guo-li Yang,Fu-ming He,Xiao-feng Yang,Xiao-xiang Wang,Shi-fang Zhao
DOI: https://doi.org/10.1016/j.tripleo.2008.03.017
2008-01-01
Oral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology
Abstract:Objective. The objective of this study was to investigate bone responses to titanium implants surface-roughened by sandblasted and double-etched treatments in a rabbit model.Study design. Sixty implants of 10 mm in length (30 machined and 30 roughened) were inserted into femurs of 30 rabbits and 30 implants of 8 mm in length (15 machined and 15 roughened) were inserted into tibias of 15 rabbits. At 2, 4, and 8 weeks postimplantation, femurs and tibias were retrieved and prepared for removal torque tests (RTQ) and histomorphometric evaluation, respectively.Results. The roughened implants showed 66.21%, 89.06%, and 115.00% greater RTQ values than did the machined implants at 2, 4, and 8 weeks. Histomorphometric evaluation demonstrated the roughened implants significantly increased bone-implant contact and peri-implant bone formation during all observation periods.Conclusion. These results suggest this surface-roughened approach provides the implant surface with a considerable osteoconductive potential promoting a high level of bone integration with bone. (Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008;106:516-24)
What problem does this paper attempt to address?