A Comparative Study Between Three-Dimensional Contrast Enhanced Ultrasound and Contrast Enhanced CT in Volumetric Measurement for Imitative Traumatic Hematoma

LIU Qiang,LUO Yukun,LV Faqin,SONG Qing,JIAO Ziyu,TANG Jie
DOI: https://doi.org/10.3969/j.issn.1005-5185.2011.03.005
2011-01-01
Abstract:Purpose To evaluate the accuracy of volumetric measurement for traumatic hematoma by three-dimensional contrast enhanced ultrasound(3D-CEUS),compared with that of contrast enhanced CT(CECT).Materials and Methods Twenty traumatic hematomas were imitated by agar gel models in different shapes and volumes.The actual volume of each model was measured by graduated cylinder,followed by 3D-CEUS and CECT volumetric measurement,and the results were compared with the actual volume.Results ①The actual volume and measured volume by 3D-CEUS and CECT were respectively 73.8±43.3ml(10.0~145.5ml),71.9±42.1ml(10.3~142.8ml),72.7±42.8ml(10.5~142.7ml).Both 3D-CEUS and CECT significantly correlated with the actual volume(r=0.99);②The measurement error using 3D-CEUS and CECT were respectively-2.4%±3.1%(-10.1%~3.7%),-1.4%±3.2%(-8.9%~5.0%)with no significant difference between the two methods(P0.05).The systematic bias and consistency scope using 3D-CEUS were-1.89ml(-5.77~2.0ml);③Good inter-rator(P0.05),intra-rator agreement(F0.05)and reproducibility were achieved with 3D-CEUS.④The mean consumed time of measurement using 3D-CEUS was 6.5±2.2min(3.2~9.5min).Conclusions 3D-CEUS performed accurate volumetric measurement for imitated traumatic hematomas with high correlation and reproducibility and no significant difference with CECT.
What problem does this paper attempt to address?