Influences of Parallel Acquisition and Conventional Acquisition Techniques on the Quality of MRI Image

GU Hai-feng,ZHENG Ling,HAO Shao-wei,LI Lin,ZHANG Zhi-qiang
DOI: https://doi.org/10.3969/j.issn.1008-8199.2012.01.008
2012-01-01
Abstract:Objective The integrated parallel acquisition technique(iPAT) is a fast acquisition technique of MRI,which has gained a wide clinical application.However,there has been no report at home addressing its influence on imaging quality.This article discusses the influences of iPAT and conventional acquisition techniques on the quality of MR images.Methods We performed scanning on a phantom and 10 volunteers using the neck matrix coil with the methods of routine/1(number of signal averaged),GRAPPA/1 and GRAPPA/2,and obtained the signal-to-noise ratio(SNR) and nonuniformity of the images through comprehensive quantitative analysis.Results With other parameters unchanged,GRAPPA/2 achieved images of better quality and higher SNR than the other two scanning methods,but GRAPPA/1 achieved the shortest scanning time.Conclusion With the scanning time unchanged,appropriate selection of average times and parallel acquisition factors can significantly improve the image quality,and with the average times unchanged,parallel acquisition techniques can significantly shorten the imaging time but somewhat reduce the image quality.
What problem does this paper attempt to address?