Continuous Monitoring of Sulfur Dioxide with A Gas Permeation Denuder-Based System

Zhong-Xian Guo,Xin-Xiang Zhang,Yi Gao,Yuan-Zong Li,Wen-Bao Chang,Yun-Xiang Ci
DOI: https://doi.org/10.1007/s00604-002-0943-3
2003-01-01
Mikrochimica Acta
Abstract:This study describes a continuous flow procedure for monitoring atmospheric sulfur dioxide using a planar gas permeation denuder (GPD) and a fiber optic spectrometer. When gaseous samples are directed through a GPD which consists of a gas-permeable membrane of poly(vinylidene) difluoride and two perspex blocks with engraved channels of mirror image, the fraction of sulfur dioxide passing the membrane is absorbed in a flowing stream of 5.0×10−4 mol L−1 5,5′-dithiobis(2,2′-dinitrobenzoic acid) (DTNB) in 0.025 mol L−1 phosphate buffer of pH 7.0, and reacts with DTNB to yield a yellow product. This product is monitored spectrophotometrically at 410 nm with a miniature charge-coupled device (CCD) fiber optic spectrometer. The analytical range of sulfur dioxide is easily adjusted via the flow rate of DTNB solution, normally from 200 ppb to 94 ppm (v/v). The procedure is hardly interfered with by coexisting metal and non-metal species, except for hydrogen sulfide in the atmosphere. The error caused by hydrogen sulfide at a sulfur dioxide level (in ppm) of 2% is less than ±10%. With the proposed procedure, the result for a mimetic air sample is very close to that of the ultraviolet fluorescence method.
What problem does this paper attempt to address?