Design of Ultrasonic Flowmeter with High Precision Based on Correlation Algorithm

YANG Zhou,HAN Yu-jie,WANG Hong,YANG Shi-yuan,FAN Shun-jie
DOI: https://doi.org/10.3969/j.issn.1001-9944.2011.05.004
2011-01-01
Abstract:Ultrasonic flowmeter based on correlation algorithm make full use of information of the ultrasonic signal,which has the advantages of high precision and strong faculty of anti-noise and so on.To inplement the correlation algorithm,much amount of calculation in the system and configurable ultrasonic signal are needed,which adds difficulty for design.The author designed an ultrasonic flowmeter with high precision based on correlation algorithm with the cooperation of Siemens corporate technology,which uses a DSP as the digital core to inplement the correlation algorithm at high speed.With neccessary analog curcuit for configurable signal and sampling,the system reaches precision of nanosecond for pipline with small caliber and low flow rate.
What problem does this paper attempt to address?