Calibration of pinhole MicroSPECT using a single point source

R Zhou,T Ma,Y Liu,Z Wu,S Wang,Y Jing,J Cheng
DOI: https://doi.org/10.1088/1748-0221/4/05/P05008
2009-01-01
Journal of Instrumentation
Abstract:To calibrate pinhole MicroSPECT system, two improved methods using only a single point source were proposed. The uniqueness of calibration result by the proposed methods was proved theoretically. And simulation calibration of a dual-head system was carried out to validate the performance of the proposed methods. The simulation results show that reasonable calibration error is given by both proposed methods.