Clinical application of ultrasound elastography hardness score and area ratio in the diffierential diagnosis of breast lesions

CAO Xiaoling,ZHANG Qunxia,LU Min,XIE Zhaopeng,ZHANG Yuxiao,PENG Jianping,ZHENG Yuanyi
DOI: https://doi.org/10.16245/j.cnki.issn1008-6978.2013.10.038
2013-01-01
Abstract:Objective To explore the clinical value of improved-5 scoring system and area ratio of ultrasound elastography in the differential diagnosis of breast lesions. Methods A total of 158 lesions underwent ultrasound elastography and hardness scoring with improved-5 scoring system. The area in the elastographic image and B-mode image were measured and the area ratio was then calculated. The pathological results of all lesions were obtained. Results The sensitivity,specificity and accuracy of improved-5 scoring system in the diagnosis of breast lesions were 93.9%、92.1%、92.4%,respectively. For area ratio method,they were 84.0% 、87.1% 、71.1%,respectively. For the combination of improved-5 scoring and area ratio,they were93.9% 、98.4% 、97.5%,respectively. Conclusion The combination of scoring and area ratio can improve the accuracy of ultrasound elastography in the differential diagnosis of breast lesions and has a high clinical value.