Application Value of Elastosonography Ratio Method and Area Ratio Method in the Identification of Benign and Malig-nant Breast Lumps

Lei YU,Jian-guo LI,Li YANG
DOI: https://doi.org/10.3969/j.issn.1007-9572.2014.24.033
2014-01-01
Chinese General Practice
Abstract:Objective To investigate the application value of elastosonography ratio method and area ratio method in the identification of benign and malignant breast lumps. Methods 45 patients with breast lump diagnosed and given operation in the department of breast surgery of Peking University Peopleˊs Hospital from December 2012 to June 2013 were selected,and there were a total of 45 lesions. Conventional ultrasound and ultrasonic elastography were used and strain ratio and area ratio were calculated. Pathology diagnosis was taken as diagnosis standard. Results Among the 45 breast masses,17 were benign and 28 were malignant. The strain ratio and area ratio between benign and malignant masses showed statistically significant difference( P<0. 05 ) . The sensitivity, specificity, best cutoff point and area under ROC curve of elastosonography ratio method were 64. 3%,82. 4%,2. 78 and 0. 77 respectively. While the sensitivity,specificity,best cutoff point and area under ROC curve of elastosonography area ratio method were 46. 4%,94. 1%,1. 95 and 0. 70. Conclusion The ultrasonic elastography shows great clinical value in the diagnosis of malignant and benign breast masses.