Effect of Ultrasound Elastography on BI-RADS in Breast Mass Diagnosis

何萍,王金锐,崔立刚,葛辉玉,雷玉涛,赵红梅,陈文,贾建文
DOI: https://doi.org/10.3760/cma.j.issn.1004-4477.2014.09.016
2014-01-01
Abstract:Objective To evaluate the effect of ultrasound elastography on the performance of BI-RADS in diagnosing breast masses.Methods An individual physician stored all the conventional ultrasound and ultrasound elastography examinations.Two sonographers performed retrospective study of all images independently.The lesions were scored with BI-RADS according to gray scale images.The lesions were scored with BI-RADS combing the ultrasound elastography after one week.Pathological diagnosis was used as the gold standard,and ROC curve was performed,to compare the diagnostic performances between the two different standards.Results ①The accuracy and specificity of the BI-RADS scoring of sonographer A was 77.90% and 74.11%,retrospectively; and the values was 88.64%,76.11% when combing with the ultrasound elastography.The accuracy and specificity of the BI-RADS scoring of sonographer B was 74.91%,71.06% without,and 83.15%,86.8% with the assistance of ultrasound elastography.Statistical difference existsed between the two scoring systems for both the sonographers.②The diagnosis of two sonographers showed medium consistency (Kappa values were 0.558 and 0.692,respectively) and no significant difference was observed between their diagnosis accuracy.Conclusions Ultrasound elastography could improve the performance of BI-RADS in breast mass diagnosis when used properly.