Determination of Hexafluoroantimonates by Ion Chromatography

JIAO Xia,LI Xiu-ling,CHEN Yong-jing,ZHU Yan
DOI: https://doi.org/10.3969/j.issn.1004-4957.2007.04.033
2007-01-01
Abstract:An ion chromatographic method was developed for the determination of hexafluoroantimonate using AG17 anion protective column and a mixture of 10 mmol/L Na_2CO_3 and 12% acetonitrile solution as eluent.It was found that the organic modifier acetonitrile,could shorten the retention time by nearly 50% and could narrow the width of the chromatographic peak by nearly 50% leading to the improvement of sensitivity and accuracy of the determination.The linear range of the calibration curve for hexafluoroantimonate was 10.0-200.0 mg/L with a correlation coefficent of 0.999 5(peak area).The detection limit was 5.0 mg/L(S/N=3).No interference was found from 5 μg/g F-,10 μg/g Cl-and NO-_3,20 μg/g SO(2-)_4 and PO(3-)_4.
What problem does this paper attempt to address?