The Performance of an L3 Hadron Calorimeter Prototype Module with BGO

C CHEN,HS CHEN,XG LI,YS LU,YF MAO,XW TANG,KL TUNG,CN WANG,KS YANG,QY YANG,HL ZHUANG,H HOFER,P LECOMTE,P LECOULTRE,Y GORODKOV,V KUTSENKO,V PLYASKIN,E SHUMILOV,V TCHUDAKOV,W BOHLEN,HT LI,B SPIESS,J ULBRICHT,SY ZHANG,T AZEMOON,RC BALL,S GOLDFARB,LW JONES,GB MILLS,BP ROE
DOI: https://doi.org/10.1016/0168-9002(88)90752-8
1988-01-01
Abstract:A prototype of a gas sampling uranium module of the L3 hadron calorimeter was built and subsequently tested with pion beams of energies between 4 and 20 GeV. The construction of the proportional chambers, the assembling of the module, the readout electronics, and the beam test are briefly described. The energy resolution of the calorimeter module alone was measured to be 30.5 ± 3% at 6 GeV, 21 ± 2% at 10 GeV and 18 ± 2% at 20 GeV. For the module together with a BGO crystal matrix in resolution was 29 ± 3% at 4 GeV, 21 ± 2% at 10 GeV, and 17 ± 2% at 20 GeV.
What problem does this paper attempt to address?