Determination of Benzhexol Hydrochloride by Capillary Zone Electrophoresis with an End-Column Electrochemiluminescence Detection

JL Yan,JF Liu,WD Cao,XH Sun,XR Yang,EK Wang
DOI: https://doi.org/10.1016/j.microc.2003.10.003
IF: 5.304
2004-01-01
Microchemical Journal
Abstract:A capillary zone electrophoresis with end-column electrochemiluminescence (ECL) detector was described for the determination of benzhexol hydrochloride. The detection was based on the tris(2,2'-bypyridine)ruthenium(II) [Ru(bpy)(3)(2+)] ECL reaction with the analyte. Electrophoresis was performed using a 25 mum i.d. uncoated capillary. 10 mM sodium phosphate buffer (pH=8.0) was used as the running buffer. The solution in the detection cell was 80 mM sodium phosphate (pH=8.0) and 5 mM)21 Ru(bpy)(3)(2+). A linear calibration curve of three-orders of magnitude was obtained (with a correlation coefficient of > 0.999) from 1.0X10(-8) to 1.0X10(-5) M and the limit of detection was 6.7 X 10(-9) M (S/N= 3). This just provides an easy and sensitive method to determine the active ingredient in pharmaceutical formulations. (C) 2003 Elsevier B.V. All rights reserved.
What problem does this paper attempt to address?