Evaluation on Adequacy of Hemodialysis by On-line Clearance Monitoring

LU Ren-hua,ZHU Ming-li,GUO Yong-ping,NI Zhao-hui,ZHANG Wei-ming,QIAN Jia-qi
DOI: https://doi.org/10.3969/j.issn.1674-8115.2006.05.025
2006-01-01
Abstract:Objective To evaluate the adequacy of hemodialysis by online clearance monitoring(OCM) in hemodialysis patients. Methods The Kt/V values of 36 maintenance hemodialysis patients in our center who were assessed by both OCM and single-pool urea kinetic model were compared,and the Kt/V values of patients in the first-use dialyser and the reused dialyser were also compared. Results In the patients who used a new dialyser,Kt/V values were 1.27±0.23 measured by OCM,whereas 1.53±0.30 by single-pool urea kinetic model(r=0.574,P0.001).In the patients who used new dialysers and reused dialysers,Kt/V values measured by OCM and single-pool urea kinetic model also showed significant differences(1.31±0.22 vs 1.55±0.30,P0.001).In the fifteen patients who applied the first,third and fifth time-used dialyser,Kt/V measured by OCM and by model had no significant differences(P0.05).In all the patients undergoing dialysis two or three times per week,Kt/V measured by OCM and by model also showed no differences(P0.05). Conclusion During hemodialysis,the OCM option provides an accurate tool for continuous online monitoring of urea clearance.
What problem does this paper attempt to address?