Assessment of stent deployment with intravascular ultrasound

LIU Ling,HUANG Cong-xin,LIN Ying-zhong,XU Guang-ma,LU Zhi-hong
DOI: https://doi.org/10.3969/j.issn.0253-4304.2006.11.011
2006-01-01
Abstract:Objective To investigate the effect of intravascular ultrasound (IVUS) in stent deployment of coronary artery disease.Methods Forty-four patients with intracoronary stent deployment were examined by intravascular ultrasound system (Endosonics).Minimal luminal diameter,cross section luminal area,total cross section,plaque area and area stenosis rate were measured before and after deployment of stent at conventional and higher pressure.Results Optimal angiographic criteria were achieved in all lesions after stenting with conventional pressure dilatation,but optimal criteria of IVUS were only obtained in four lesions.Forty lesions failed to achieve optimal criteria of IVUS and 35 lesions were treated with post-1dilatation with 14atm to 16atm,then twenty-two lesions(50) obtained the optimal criteria of IVUS.Cross section luminal area increased from (7.3±1.7)mm~2 to (9.2±1.5)mm~2 after higher pressure dilatation(P<0.001).Thare was significant difference in the minimal luminal diameter,minimnl cross section luminal area,total cross section,plaque area and area stenosis rate before and after stent deployment.Conclusions IVUS is an important and useful tool in assessment of stent deployment with coronary artery disease.All stent deployment under conventional pressure inflation could not achieve optimal criteria of IVUS.And post-dilatation with IVUS guidance could improve effects of intervention.