Application of Intravascular Ultrasound Imaging in Stent-grafting of Aortic Dissecting Aneurysm

蒋俊豪,符伟国,郭大乔,陈斌,杨珏,王玉琦
DOI: https://doi.org/10.3969/j.issn.0253-9934.2003.08.006
2003-01-01
Abstract:Objective To assess the role of intravascular ultrasound (IVUS) imaging in the stent grafting of dissecting aortic aneurysm. Methods Between April 2001 and December 2002, 7 patients with dissecting aneurysm of the descending thoracic aorta received stent grafting under the guidance of IVUS. The role of IVUS was assessed in the size of the stent graft'selection, the deployment site and observation of the therapeutic effect. Results In the stent grafting of dissecting aneurysm of the descending thoracic aorta, stent graft size could be determined by the maximum diameter of the proximal true lumen measured by IVUS. The deployment location could be demonstrated by measuring the distance between the entry site and the orifice of the left subclavian artery during IVUS. The IVUS monitored snare technique could ensure the stent graft to be deployed in the true lumen. With the aid of IVUS, type I endoleak could be detected in time by observing the anchoring of the stent graft to the true lumen. Conclusion In the stent grafting of dissecting aneurysm of the descending thoracic aorta, IVUS can provide the guidance for confirmation of the stent graft size, the deployment location and evaluation of the therapeutic effect.