Double-Beam Laser Indirect Absorption Detection in Capillary Electrophoresis

YJ XUE,ES YEUNG
DOI: https://doi.org/10.1021/ac00068a034
IF: 7.4
1993-01-01
Analytical Chemistry
Abstract:The use of a laser for optical detection in capillary electrophoresis (CE) allows efficient light coupling. By increasing the absorption path length and by reducing noise through an all-electronic noise canceller, the performance of indirect absorption detection is substantially enhanced. In 75-mum capillaries, as low as 1 X 10(-7) M pyruvate (10(-15) mol injected) can be detected. In 14-mum capillaries, 3 x 10(-6) M (1.5 x 10(-16) mol injected) can be detected. When a cationic chromophore is used, K+ at 1 x 10(-6) M injected is detectable. These represent the best performance to date, by over 1 order of magnitude, for indirect absorption detection in CE without preconcentration.