Optimization Of Capillary Electrophoresis-Laser Induced Fluorescence Detection With Sheath-Flow Cuvette And Its Application To Genetic Analysis

Lin Chen,Jicun Ren
DOI: https://doi.org/10.3321/j.issn:0253-3820.2003.12.002
IF: 1.193
2003-01-01
Chinese Journal of Analytical Chemistry
Abstract:We systematically investigated the effects of some factors, such as the sheath-flow velocity, laser power and detection position, on capillary electrophoresis-laser induced fluorescence detection (CE-LIF) with a sheath-flow cuvette. The data demonstrated that the detection signals were markedly decreasing, but the numbers of theoretical plate were significantly increasing with an increase in the sheath-flow. In the low laser power, the detection signals linearly increased with laser power. Under the optimum conditions, CE-LIF has a wide linear range from 1x10(-8) to 1x10(-10) mol/L, and a low detection limit of 7.5x10(-11) mol/L (S/N=3) for fluorescein. This system was. successfully used for the detection of deoxyribonucleic acid (DNA) fragments and polymerase chain reaction (PCR) products using polydimethylacrylamide as a sieving medium.
What problem does this paper attempt to address?