Method Improvement on the Turbidity Point Determination of Iron Sucrose Injection

LI Li,HUA Chuan-zheng,FANG Wen
DOI: https://doi.org/10.3969/j.issn.1673-7806.2013.03.012
2013-01-01
Abstract:Objective:To establish an objective method for the evaluation of turbidity point in an iron sucrose injection.Methods:In a certain concentration of sample solution,the 0.1 mol.L-1 hydrochloric acid solution was continuously added and stirred,the solution was monitored by a turbidity meter until the turbidity reached the range of 0.2~0.8 NTU,the stirring was then stopped and the pH value of the solution was recorded,this pH value was the sample turbidity point.Results:The method applied instruments to test the iron sucrose injection turbidity points,the relative standard deviation of precision test was less than 1.0%.Conclusion:This method can effectively avoid the subjective interfering factors on the determination of the turbidity point,so it is accurate and reliable.
What problem does this paper attempt to address?