Evaluation of YZ54 A Scan in Axial Length Measurement of Pre-operative Cataract Surgery

JIANG Yong-xiang,LU Yi,JIANG Yong-ming
DOI: https://doi.org/10.3969/j.issn.1671-2420.2007.01.016
2007-01-01
Abstract:Purpose To evaluate the accuracy and safety-of YZ54 A Scan in axial length measurement of pre- operative cataract surgery.Methods The axial lengths of 84 eyes(84 cases)were measured with both NIDEK US-1800 A Scan(NIDEK A Scan group)and YZ54 A Scan(YZ54 A Scan group).Results The axial lengths of all the eyes were measured successfully without any complications.Based on the axial lengths measured,84 eyes were divided into three groups:short axial length group(axial length was less than 23 mm;23 eyes),normal axial length group(axial length was equal to or more than 23 mm and equal to or less than 25 mm;27 eyes)and long axial length group(axial length was more than 25 mm;34 eyes).There were no statistically significant differences in axial length,the depth of central anterior chamber,the depth of lens and the length of vitreous cavity between NIDEK A Scan group and YZ54 A Scan group.Conclusion The accuracy and safety of YZ54 A Scan in axial length measurement of pre-operative cataract surgery are eneouragble which are as the same as that of NIDEK US-1800 A Scan.This kind of Scan meets the chnical acquirement.(Chin J Ophthalmol and Otorhinolaryngol,2007,7:39-40)