Determination of Cholesterol in Human Serum by Flow Injection Chemiluminescence Method

WANG Jing-wu,WU Dong-mei,YANG Jia
DOI: https://doi.org/10.3969/j.issn.1004-1656.2007.07.015
2007-01-01
Abstract:Hemoglobin was used as a mimetic peroxidase in the catalytic chemiluminescence reaction of lumino and hydrogen peroxide. The hemoglobin-catalyzed reaction was coupled with the oxidation reaction of cholesterol catalyzed by COA(cholesterol oxydase , COA.) to determine total cholesterol(TC)and free cholesterol (FC) in human serum by flow injection chemiluminescence method. The sampling frequency was 30 s/h. The linear range for free cholesterol is 5.6×10-7mol/L~3.8×10-5mol/L and the detection limit is 3.0×10-7mol/L. The linear range for total cholesterol is 1.1×10-6mol/L~4.3×10-5mol/L and the detection limit is 1.5×10-7mol/L.
What problem does this paper attempt to address?