Evaluation of the Tube Diffusion Test Detectingβ-Lactam Residue in Raw Milk

WU Xia,SHAO Hui,WANG Xin,ZHANG Lanwei
DOI: https://doi.org/10.3969/j.issn.1005-9369.2011.05.006
2011-01-01
Abstract:The principle of the tube diffusion test was based on the method of microbiological inhibitor.The bacteria of Bs was used as the test bacteria.And the indicator was added into the test medium.Six kinds of β-lactam antibiotics were tested at five concentrations based on estab-lished codex alimentarius maximum residue limits(MRLs).Test parameters studied were compared to the method of Delvotest SP.The agreement between the methods of tube diffusion and Delvotest SP was high.The limits of detection(LODs) of the six kinds of β-lactam antibiotics were1.0×MRL.The sensitivity and the specificity at 1.0×MRL was 97% and 91% respectively.And the detecting time was 3 h and 10min.The results of stability of experiment indicated the test tube prepared in the experiment could be stored at 4 ℃and 15 ℃ for 7days and at room tem-perature for 3 days,which would not affect the testing results.Thus the method of tube diffusion could be tried to developed a test kit.