Determination of β-lactam antibiotics in milk using micro-flow chemiluminescence system with on-line solid phase extraction

Wei Liu,Zhujun Zhang,Zuoqin Liu
DOI: https://doi.org/10.1016/j.aca.2007.04.027
IF: 6.911
2007-01-01
Analytica Chimica Acta
Abstract:In this paper, a chemiluminescence (CL) micro-flow system combined with on-line solid phase extraction (SPE) is presented for determination of β-lactam antibiotics (penicillin, cefradine, cefadroxil, cefalexin) in milk. It is based on the enhancement effect of β-lactam antibiotics on the luminol-K3Fe(CN)6 CL system. The micro-flow system was fabricated from two polymethyl methacrylate (PMMA) plates (50mm×40mm×5mm) with the microchannels of 200μm wide and 150μm deep. C18-modified silica gel was packed into the microchannel (length: 10mm; width: 1mm; depth: 500μm) to serve as SPE device. Extraction and preconcentration of the analytes were carried out using on-line SPE micro-flow system and the selectivity of CL detection was improved. The detection limits were 0.5μgmL−1 of penicillin, 0.04μgmL−1 of cefradine, 0.08μgmL−1 of cefadroxil and 0.1μgmL−1 of cefalexin. The proposed method was also applied to analyze the β-lactam antibiotics in milk. Experimental results were in good agreement with those obtained by high performance liquid chromatography (HPLC) method with UV detection.
What problem does this paper attempt to address?