Purge and trap/gas chromatography-mass spectrometry method for the determination of 8 chlorinated hydrocarbons in source water

MA Jun-hua,ZHANG Hai-yan,ZHANG Cheng-qiang
2012-01-01
Abstract:Objective:To investigate the purge-and-trap/ gas chromatography-mass spectrometry method for the determination of 8 chlorinated hydrocarbons in source water.Methods: Target organic pollutants in source water were collected by purge and trap,then analyzed using gas chromatography-mass spectrometry.The qualitation was accomplished with use of the target component spectra and their retention time,and the quantitation was accomplished by use of characteristic ions.Results: The recoveries were 91%~118% with the relative standard deviation of 1.19% ~4.69%,the detection limit was 0.026 μg/L~0.163 μg/L.Conclusion: The method is easy,sensitive,accurate,which can be used for determination of 8 chlorinated hydrocarbons in water samples.
What problem does this paper attempt to address?