Chemiluminescence Flow Biosensor for Hydrogen Peroxide with Immobilized Reagents

BX Li,W Qin,ZJ Zhang
DOI: https://doi.org/10.1016/s0925-4005(00)00623-7
IF: 6.911
1998-01-01
Analytica Chimica Acta
Abstract:A novel chemiluminescence (CL) sensing system for hydrogen peroxide combined with flow-injection analysis (FIA) is described in this paper. It is prepared by electrostatically immobilizing the analytical reagents, luminol and cobalt(II), on a strongly basic anion-exchange resin and a weakly acid cation-exchange resin, respectively. Hydrogen peroxide is sensed by the CL reaction with luminol and cobalt(II) bleeding from the ion-exchange column with immobilized reagents by hydrolysis. The calibration graph is linear in the range 4×10−8 to 1×10−5moll−1, and the detection limit is 1.2×10−8moll−1 hydrogen peroxide. A complete analysis could be performed in 1min with a relative standard deviation of <5%. The system could be used for 50h and has been applied successfully to the determination of hydrogen peroxide in rainwater and glucose in serum by measuring the formation of hydrogen peroxide from a packed bed reactor with immobilized glucose oxidase.
What problem does this paper attempt to address?